Chapter 3 Calibration Procedures
Aborting a Calibration in Progress
68
19
Pressing the "Calibrate" key saves the new calibration constants for
the output current and goes to the OCP calibration mode.
A "CALIBRATING" message appears for one second to indicate that the current
calibration is progressing and new calibration constants of "SETUP 3" are
stored. Then, the display shows the above message to indicate that the power
supply is ready for the OCP calibration.
If the calibration fails, an "DAC CAL FAIL" or "ADC CAL FAIL" message appears
for one second and the display shows the "CAL SETUP 3" for current
calibration again.
20
Pressing the "Calibrate" key saves the new OCP calibration constants
and return to the calibration mode.
A "CALIBRATING" message appears for several seconds to indicate that the
OCP calibration is progressing and new OCP calibration constants of "SETUP
4" are stored. Then the display will return to the calibration mode.
21
Turn off the power supply to exit the calibration mode.
Aborting a Calibration in Progress
Sometimes it may be necessary to abort a calibration after the procedure has
already been initiated. You can abort a calibration at any time by turning the
power supply off from the front panel. When performing a calibration from the
remote interface, you can abort a calibration by issuing a remote interface
device clear message or by pressing the front-panel "Local" key.
cal setup 4
cal mode
Calibrate
Calibrate
Power