0

Out of 0 Ratings

Owner's of the Agilent Technologies Video Gaming Accessories Agilent Technologies Video Gaming Accessories gave it a score of 0 out of 5. Here's how the scores stacked up:
  • Reliability

    0 out of 5
  • Durability

    0 out of 5
  • Maintenance

    0 out of 5
  • Performance

    0 out of 5
  • Ease of Use

    0 out of 5
of 128
 
Chapter 3 Calibration Procedures
General Calibration/Adjustment Procedure
67
3
16
Read the DVM and change the middle current value on the display to
match the computer current (DVM reading
¸ by shunt resistance).
For example, if the computered value is 3.499 A, adjust the current to 3.499 A
using the knob and arrow keys.
Notice that you should wait for the DVM reading to stabilize for accurate
calibration.
17
Pressing the "Calibrate'' key saves the change and selects the high
current calibration point.
If the entered number is within an acceptable range, an "ENTERED" message
appears for one second. If the entered number is not correct, an "MIN VALUE"
or "MAX VALUE" message appears for one second and the display shows the
middle current calibration point again. The display now shows th e high current
calibration point
18
Read the DVM and change the high current value on the display to
match the computed current (DVM reading
¸ by shunt resistance).
For example, if the computed value is 6.899 A, adjust the current to 6.899 A
using the knob and arrow keys.
Notice that you should wait for the DVM reading to stabilize for accurate
calibration.
i mi 3.499 a
i hi 6.900 a
i hi 6.899 a
Calibrate