0

Out of 0 Ratings

Owner's of the Agilent Technologies Video Gaming Accessories Agilent Technologies Video Gaming Accessories gave it a score of 0 out of 5. Here's how the scores stacked up:
  • Reliability

    0 out of 5
  • Durability

    0 out of 5
  • Maintenance

    0 out of 5
  • Performance

    0 out of 5
  • Ease of Use

    0 out of 5
of 128
 
Chapter 3 Calibration Procedures
Constant Voltage (CV) Verifications
49
3
Constant Voltage (CV) Verifications
Constant Voltage Test Setup
If more than one meter or a meter and an oscilloscope are used, connect each
to the (+) and (-) terminals by a separate pair of leads to avoid mutual coupling
effects. Use coaxial cable or shielded 2-wire cable to avoid noise pick-up on
the test leads.
Voltage Programming and Readback Accuracy
This test verifies that the voltage programming and GPIB or RS-232 readback
functions are within specifications. Note that the readback values over the
remote interface should be identical to those displayed on the front panel.
You should program the power supply over the remote interface for this test
to avoid round off errors.
1
Turn off the power supply and connect a digital voltmeter between the (+) and
(-) terminals of the output to be tested as shown in Figure 3-1.
2
Turn on the power supply. Select the 30V/4A range and enable the outputs by
sending the commands:
VOLT:RANG P30V
OUTP ON
3
Program the output voltage to zero volt and current to full rated value (4.0A)
by sending the commands:
VOLT 0
CURR 4
4
Record the output voltage reading on the digital voltmeter (DVM). The reading
should be within the limits of (0V
±
10 mV). Also, note that the
CV, Adrs, Limit,
and
Rmt annunciators are on.
5
Readback the output voltage over the remote interface by sending the
command:
MEAS:VOLT?
6
Record the value displayed on the controller. This value should be within the
limits of (DVM
±
5 mV)