Chapter 3 Calibration Procedures
Performance Test Record for Agilent E3632A
57
3
Performance Test Record for Agilent E3632A
CV Performance Test Record
CC Performance Test Record
Test Description
Actual
Result
Specifications
Upper Limit Lower Limit
CV Programming Accuracy @ 0 volts
(DVM reading)
+0.0100 V -0.0100 V
CV Readback Accuracy @ 0 volts DVM + 0.0050 V DVM - 0.0050 V
CV Programming Accuracy @ Full Scale
(DVM reading)
+30.025V 29.9750 V
CV Readback Accuracy @ Full Scale DVM + 0.0200 V DVM - 0.0200 V
CV Load Regulation Maximum change: <5 mV
CV Line Regulation Maximum change: <5 mV
CV Ripple/Noise <2 mV p-p, 0.35 mV rms
Load Transient Response Time <50
msec
Test Description
Actual
Result
Specifications
Upper Limit Lower Limit
CC Programming Accuracy @ 0 amps (I
O
) +0.0100 A -0.0100 A
CC Readback Accuracy @ 0 amps I
O
+ 0.0050 A I
O
- 0.0050 A
CC Programming Accuracy @ Full Scale (I
O
) 7.0240 A 6.9760 A
CC Readback Accuracy @ Full Scale I
O
+ 0.0155 A I
O
- 0.0155 A
CC Load Regulation Maximum change: <0.95 mA
CC Line Regulation Maximum change: <0.95 mA
CC Ripple/Noise <2 mA
Common Mode Current Noise <1.5
mA rms