24
Press READ/TIMER to start the average reading.
The display will show blinking dashes and
the icons for cuvet, detectors and lamp will
appear during measurement. When a new
intermediate value is displayed, the cuvet icon
and the measurement unit will briefly blink.
When the measurement is ended, the final
averaged result is displayed directly in NTU.
RANGE AND UNITS
HI 93414 automatically selects the correct turbidity
range to display the results with the highest accuracy.
If the measured value is higher than 1000 NTU
(over range), the display will show the maximum
value blinking.
The instrument has an EPA compliance reading
mode. If this feature is activated in SETUP, “EPA”
tag will appear on the LCD and the readings will be
rounded to meet EPA reporting requirements as
shown in the table.
UTN
t
seraeNotdroceR
0.1-0.0 50.0
01-1 1.0
04-01 1
001-04 5
004-001 01
0001-004 05
0001> 001