0

Out of 0 Ratings

Owner's of the Hanna Instruments Marine Instruments Turbidity and Free/Total Chlorine Meter gave it a score of 0 out of 5. Here's how the scores stacked up:
  • Reliability

    0 out of 5
  • Durability

    0 out of 5
  • Maintenance

    0 out of 5
  • Performance

    0 out of 5
  • Ease of Use

    0 out of 5
of 52
 
In order to match multiple cuvets follow the next
steps:
Fill some cuvets with high quality water
(<0.1NTU) up to the 10 mL mark.
Clean and oil the cuvets as described before.
Turn the instrument ON.
Insert the first cuvet into the instrument and
press READ/TIMER
. Record the reading.
Record the position of the cuvet and the
displayed reading.
Mark this position on the thicker white band on
the top of the cuvet with a water resistant pencil.
Insert the second cuvet into the instrument and
take a reading.
MATCHING MULTIPLE CUVETS
Precise measurements require the use of a single cuvet. If it is not possible, cuvet selection and
matching must be performed before taking measurements.