Chapter 3 Calibration Procedures
Calibration Security Code
63
3
4
Unsecure the power supply.
The power supply is unsecured when you press the key. You will see
the above message from the front panel for one second. The "CAL MODE"
message is displayed on the front panel after above message.
5
Turn off the calibration mode.
Turn off the power supply to exit the calibration mode.
unsecured
To re-secure the power supply (following calibration), perform this
procedure again.
Secure
Secure
Power