0

Out of 0 Ratings

Owner's of the Agilent Technologies Welding System Agilent Technologies Welding System gave it a score of 0 out of 5. Here's how the scores stacked up:
  • Reliability

    0 out of 5
  • Durability

    0 out of 5
  • Maintenance

    0 out of 5
  • Performance

    0 out of 5
  • Ease of Use

    0 out of 5
of 134
 
Chapter 1
Specifications
Supplemental Characteristics
17
1
Supplemental Characteristics
Output Programming Range (maximum programmable values)
+6V Output +25V Output -25V Output
Voltage 0 to 6.18 V 0 to 25.75 V 0 to -25.75 V
Current 0 to 5.15 A 0 to 1.03 A 0 to 1.03 A
Temperature Coefficient, ±(% of output + offset)
Maximum change in output/readback per °C after a 30-minute warm-up
+6V Output +25V Output -25V Output
Voltage 0.01% + 2 mV 0.01% + 3 mV 0.01% + 3 mV
Current 0.02% + 3 mA 0.02% + 0.5 mA 0.02% + 0.5 mA
Stability, ±(% of output + offset)
Following a 30-minute warm-up, change in output over 8 hours under constant load,
line, and ambient temperature
+6V Output +25V Output -25V Output
Voltage 0.03% + 1 mV 0.02% + 2 mV 0.02% + 2 mV
Current 0.1% + 3 mA 0.05% + 1 mA 0.05% + 1 mA
Voltage Programming Speed
Maximum time required for output voltage to settle within 1% of its total excursion (for
resistive load). Excludes command processing time.
+6V Output +25V Output -25V Output
Full load Up 11 msec 50 msec 50 msec
Full load Down 13 msec 45 msec 45 msec
No load Up 10 msec 20 msec 20 msec
No load Down 200 msec 400 msec 400 msec
Isolation
The 0-6V supply is isolated from the ±25V supply up to ±240 Vdc. Maximum isolation
voltage from any terminal to chassis ground ±240 Vdc.
AC Input Ratings (selectable via rear panel selector)
std 115 Vac ± 10%, 47 to 63 Hz
opt 0E3 230 Vac ± 10%, 47 to 63 Hz
opt 0E9 100 Vac ± 10%, 47 to 63 Hz
Maximum Input Power
350 VA with full load