of 134
 
Chapter 1 Specifications
Performance Specifications
15
1
Performance Specifications
Output Ratings (@ 0°C - 40°C)
+6V Output 0 to +6 V ; 0 to 5 A
+25V Output 0 to +25 V ; 0 to 1 A
-25V Output 0 to -25 V ; 0 to 1 A
Programming Accuracy
[1]
12 months (@ 25°C ±
5°C), ±(% of output + offset)
+6V Output
+25V Output -25V Output
Voltage 0.1% + 5 mV 0.05% + 20 mV 0.05% + 20 mV
Current 0.2% + 10 mA 0.15% + 4 mA 0.15% + 4 mA
Readback Accuracy
[1]
12 months (over GPIB and RS-232 or front panel with respect
to actual output @ 25°C ± 5°C), ±(% of output + offset)
+6V Output
+25V Output -25V Output
Voltage 0.1% + 5 mV 0.05% + 10 mV 0.05% + 10 mV
Current 0.2% + 10 mA 0.15% + 4 mA 0.15% + 4 mA
Ripple and Noise (with outputs ungrounded, or with either output terminal grounded,
20 Hz to 20 MHz)
+6V Output
+25V Output -25V Output
Voltage <0.35 mV rms <0.35 mV rms <0.35 mV rms
<2 mV p-p <2 mV p-p <2 mV p-p
Current <2 mA rms <500 mA rms <500 mA rms
Common mode current <1.5 mA rms
Load Regulation,
±(% of output + offset)
Change in output voltage or current for any load change within ratings
Voltage <0.01% + 2 mV
Current <0.01% + 250 mA
Line Regulation,
±(% of output + offset)
Change in output voltage and current for any line change within ratings
Voltage <0.01% + 2 mV
Current <0.01% + 250 mA
[1]
Accuracy specifications are after an 1-hour warm-up and calibration at 25°C