Chapter 1 Specifications
Performance Specifications
16
Programming Resolution
+6V Output
+25V Output -25V Output
Voltage 0.5 mV 1.5 mV 1.5 mV
Current 0.5 mA 0.1 mA 0.1 mA
Readback Resolution
+6V Output
+25V Output -25V Output
Voltage 0.5 mV 1.5 mV 1.5 mV
Current 0.5 mA 0.1 mA 0.1 mA
Meter Resolution
+6V Output
+25V Output -25V Output
Voltage 1 mV 10 mV 10 mV
Current 1 mA 1 mA 1 mA
Transient Response Time
Less than 50 msec for output recover to within 15 mV following a change in output
current from full load to half load or vice versa
Command Processing Time
Programming Commands : Maximum time for output to change after receipt of APPLy
and SOURce commands : <50 msec
Readback Command : Maximum time to readback output by MEASure? command :
<100 msec
The Other Commands : < 50 msec
Tracking Accuracy
The
±25V outputs track each other within ± (0.2% of output + 20 mV).