of 136
 
Chapter 1 Specifications
Supplemental Characteristics
17
1
Supplemental Characteristics
Table 1-2. Supplemental Characteristics
Remote Sensing Capability
Voltage drop Up to 0.7 V per each lead
Load regulation Add 5 mV to spec for each 1-volt change in the + output
lead due to load current changes.
Load voltage Subtract voltage drop in load leads from specified output
voltage rating.
Temperature Coefficient, ±(% of output + offset)
Maximum change in output/readback per °C after a 30-minute warm-up
Voltage 0.01% + 3 mV
Current 0.02% + 3 mA
Stability, ±(% of output + offset)
Following 1 hour warm-up, change in output over 8 hours under constant load,
line, and ambient temperature
Voltage 0.02% + 1 mV
Current 0.1% + 1 mA
Parameter Agilent E3633A Agilent E3634A
Output Programming Range
(maximum programmable values)
Low Range 0 to +8.24 V/
0 to 20.6 A
0 to +25.75 V/
0 to 7.21 A
High Range 0 to +20.6 V/
0 to 10.3 A
0 to +51.5V/
0 to 4.12 A
OVP 1 V to 22 V 1 V to 55 V
OCP 0 A to 22 A 0 A to 7.5 A
Voltage Programming
Speed: Maximum time required
for output voltage to settle within
1% of its total excursion (for
resistive load). Excludes
command processing time.
Full Load No Load Full Load No Load
Up 95 msec 45 msec 80 msec 100 msec
Down 30 msec 450 msec 30 msec 450 msec