of 136
 
44
Calibration Procedures
This chapter contains procedures for verification of the power supply’s
performance and calibration (adjustment). The chapter is divided into the
following sections:
Agilent Technologies Calibration Services, on page 45
Calibration Interval, on page 45
Automating Calibration Procedures, on page 46
Test Considerations, on page 46
Recommended Test Equipment, on page 47
Performance Verification Tests, on page 48
Measurement Techniques, starting on page 49
Constant Voltage (CV) Verifications, starting on page 51
Constant Current (CC) Verifications, starting on page 56
Common Mode Current Noise, on page 59
Performance Test Record for Agilent E3633A/E3634A, on page 60
Calibration Security Code, starting on page 61
Calibration Count, on page 64
Calibration Message, on page 64
General Calibration/Adjustment Procedure, starting on page 65
Aborting a Calibration in Progress, on page 71
Calibration Record for Agilent E3633A/E3634A, on page 72
Error Messages, starting on page 73
An Example Program of Excel 97 for Calibration, starting on page 75
The performance verification tests for constant voltage (CV) and constant
current (CC) operations use the power supply’s specifications listed in
chapter 1, ‘‘Specifications,’’ starting on page 13.