0

Out of 0 Ratings

Owner's of the 3M Pet Fence Littmann Electronic Stethoscope gave it a score of 0 out of 5. Here's how the scores stacked up:
  • Reliability

    0 out of 5
  • Durability

    0 out of 5
  • Maintenance

    0 out of 5
  • Performance

    0 out of 5
  • Ease of Use

    0 out of 5
of 15
 
10
GB English
11
GB English
To select auto off power management: Select the AUTO OFF option in the SETUP menu. Use the (-) and (+) buttons to scroll
through the list of options. Press and release the (M) button to select. The factory default is ENABLE.
Two different settings work together to infl uence the length of time the Model 3200 is on full power: The AUTO OFF setting
(ENABLE vs. DISABLE) and the POWER setting (HIGH, MEDIUM, and LOW - HIGH setting uses more battery power, LOW setting
uses less battery power). To increase battery life, ENABLE the AUTO OFF feature and select the LOW power setting.
When the AUTO OFF feature is enabled, the stethoscope will enter standby mode after a period of time determined by the
POWER setting if there is no contact detected on the diaphragm AND the buttons are not activated. When the AUTO OFF
feature is disabled, the stethoscope will enter standby mode after a period of time determined by the POWER setting from
the last button activation. In addition, when the AUTO OFF feature is disabled, there will be an audible alert 10 seconds prior
to entering standby mode. After entering standby mode, the Model 3200 will be powered on when contact is detected on the
diaphragm or by button activation. Otherwise, the Model 3200 will completely power off after the period of time determined
by the POWER setting (see below).
With AUTO OFF feature ENABLED:
SETUP Options Detail
POWER HIGH
Remains on full power for 30 seconds before entering standby
Remains in standby for 5 hours before powering off
Backlight remains lit for 5 seconds after pressing the power button
MEDIUM
Remains on full power for 20 seconds before entering standby
Remains in standby for 2 hours before powering off
Backlight remains lit for 5 seconds after pressing the power button
LOW
Remains on full power for 10 seconds before entering standby
Remains in standby for 30 minutes before powering off
Backlight remains lit for 3 seconds after pressing power button
To specify the preset (default) volume level: Select the SET VOL option in the SETUP menu. Press and release the (M)
button to select the current volume level as default. The factory default is level 3.
To specify the preset (default) fi lter setting: Select the SET FLTR option in the SETUP menu. Press and release the (M)
button to select the current fi lter as default. The factory default is the diaphragm fi lter.
To select the backlight setting: Select the BACKLITE option in the SETUP menu. Use the (-) and (+) buttons to scroll
through the list of options. Press and release the (M) button to select. The factory default is ALWAYS. The ALWAYS option will
illuminate the backlight at all times at a reduced level while the stethoscope is powered on and not in standby mode. The ON
DMND option will illuminate the backlight for fi ve seconds by depressing and releasing the power button. The DISABLE option
disables the backlight at all times and will conserve power.